Case Report - Interventional Cardiology (2018) Volume 10, Issue 1

Successful retrieval of a dislodged Zotarolimus-eluting coronary stent in the ascending aorta

Corresponding Author:
Adithya Udupa
Apollo Hospital, Mysore, Karnataka, India
Tel: +7506242952 E-mail: udupaputti@gmail.com

Submitted: 10 June 2017; Accepted: 29 January 2018; Published online: 05 February 2018

Abstract

Coronary stent dislodgment or embolization is rare in the era of 3rd generation Drug Eluting Stents but is a challenging complication in interventional cardiology. Intracoronary embolization of the dislodged stent is associated with a high risk of coronary occlusion, due to thrombus formation and subsequent myocardial infarction. Furthermore, systemic embolization may cause severe cerebrovascular events. Nonsurgical retrieval options for this complication are being used with good results in most cases, but bailout cardiac surgery may be indicated if percutaneous retrieval attempts fail. We present here a case of dislodgement of a zotarolimus-eluting coronary stent during percutaneous coronary intervention and successful retrieval from the ascending aorta using a Amplatz goose neck microsnare.

Keywords

Coronary stent dislodgment; Zotarolimus-eluting coronary stent; Amplatz goose neck microsnare and Percutaneous coronary intervention

Introduction

Coronary stent dislodgement is a rare but known complication of percutaneous coronary intervention [1]. Stent loss has been associated with severe complications, including embolization to the coronary circulation, embolic cerebrovascular events and peripheral embolization. Retrieval of a dislodged stent can be done either percutaneously or surgically [2]. We report a rare case of stent dislodgement in the ascending thoracic aorta during percutaneous coronary intervention to a calcified proximal left circumflex artery (LCX) lesion.

Case Report

A 78-year-old hypertensive and diabetic gentleman was admitted to our hospital with a history of exertional angina since 3 months and past history of percutaneous coronary intervention (PCI) to LAD 4 years back. Coronary angiography showed calcified major coronaries, patent LAD stent with 80% diffuse lesion of proximal left circumflex artery extending into Obtuse marginal (Figure 1a) and normal RCA. Patient was taken up for PCI to LCX. 6Fr XB (Xtra backup/3.5, Cordis Corporation) guiding catheter was used to engage the left main via right femoral approach. The lesion was crossed with a 0.014” Fielder FC (ASAHI) guide wire. Lesion was serially dilated with a 1.25×10, 1.5×10, 2.0×10 balloons (Figure 1b). After what was thought to be an adequate dilatation, a 2.5×30 mm zotarolimus-eluting coronary stent (Resolute integrity, Medtronic, Inc) was tried to pass across the lesion. There was considerable resistance for the stent delivery. The lesions were again dilated with a 2.5×10 mm balloon and another 0.014” guiding wire was used as support. However, again stent had significant resistance to cross through the lesion (Figure 2a). The stent was again retracted and rotablation was planned to have adequate bed preparation. However on retracting the stent catheter, we were surprised to see only the balloon without stent. A search for the dislodged stent was made under fluoro guidance. As the dislodged stent could not be found in the coronaries or anywhere on the guidewire it was thought to be somewhere inside the guiding catheter. Hence the whole system was removed insitu and a search was made inside the guiding catheter. But however there were no signs of the dislodged stent inside the guiding catheter or guiding wire.

interventional-cardiology-caudal-lesion

Figure 1: a) RAO caudal view showing diffuse calcified lesion of LCX and OM. b) showing predilation of the lesion in same view.

interventional-cardiology-floating-edge

Figure 2: a) RAO caudal view showing guide catheter buckling up on attempt to stent deployment. b) Showing dislodged stent freely floating in ascending aorta in AP view, vertical arrow head showing fixation of proximal end to LCX ostia and horizontal arrow head pointing the distal floating edge.

Again a renewed effort was made to search for the dislodged stent under double magnification in various angles. Now we could see the stent hanging precariously, unwound extending from the proximal LCX and into the ascending aorta (Figure 2b). An Amplatz Goose neck Microsnare was pased through the XB guiding catheter and an attempt to grasp the stent was made, unsuccessfully. Another attempt with same snare was tried using a Judkins right guiding catheter. Now the snare was able to grasp the floating stent (Figure 3a) and with a mild tug the stent was freed from the calcified LCX and was able to be retrieved (Figure 3b). Patient remained hemodynamically stable throughout the procedure.

interventional-cardiology-photograph-snare

Figure 3: a) showing floating edge being captured by the snare and retrieving into JR guiding (arrow head), b) the photograph of retrieved stent along with the snare.

Discussion

Stent dislodgement is a rare complication that has been reported in 0.32–8.4% of PCI procedures [1]. The incidence of stent loss has been decreasing in recent times primarily due to technological advances in PCI and stent design. Brilakis et al., [3] reported that the annual incidence of stent loss during PCI in a single center, increased from 1994 to 1997, but after that subsequently declined. However, stent dislodgement is still a potential cause of serious complications such as death, stroke, embolization, and need for an emergency operation [3].

Dislodgement of a stent can be secondary to extreme coronary angulation, highly calcified coronary arteries, inadequate bed preparation, and direct stenting [4]. Thus, adequate bed preparation with predilation or rotablation may help prevent stent loss. In our case, the stent dislodgement was most probably secondary to the highly calcified LCX, angulation and possible inadequate coronary artery predilation. A lower threshold for rotablation could have possibly prevented this complication.

Numbers of techniques have been described for retrieval of dislodged coronary stent like, a small balloon technique, a double wire technique, or a loop snare. The choice of retrieval device and technique usually depends on the location of the stent, the type of stent, its deployment status and operator experience. A snare loop is relatively safe and easy to use. It has a low rate of complications and appears to be effective [5].A snare loop appeared to be the best method of retrieval in our case, as free edge of lost stent was hanging in the ascending thoracic aorta.

Nowadays, stent loss and dislodgement occur rarely. However, every catheterization laboratory should be equipped with instruments for intravascular foreign body retrieval and interventional cardiologists should be familiar with these retrieval methods and techniques.

References

Awards Nomination 20+ Million Readerbase

Select your language of interest to view the total content in your interested language

Journal Metrics:

Impact Factor 1.34
Scimago Journal Rank (SJR) 123
SJR Total Cites 15
Source Normalized Impact per Paper (SNIP) 0.144
h-index (2023) 12
PubMed NLM ID:  10148499
Google Scholar h5 index: 6
Iindex Copernicus Value: 105.52


Google Scholar citation report
Citations : 1297

Interventional Cardiology received 1297 citations as per Google Scholar report


Interventional Cardiology peer review process verified at publons

Indexed In

flyer